Minimum Order Quantity | 1 Piece |
Rhythm record | record setup according to rhythm record format and mode, automatically measuring and analysing. |
Manual record | record setup according to manual record format, manually changing leads. |
Input impedance | >= 50M |
Frequency response | 0.05Hz~150Hz(-3dB~+0.4dB) |
Recording way | Thermal printing system |
Specification of recording paper | 210mm (W)x20m(L) high-speed thermal paper |
Sensitivity selections | 2.5,5,10,20,40mm/mV, error: +5%. Standard sensitivity is 10mm/mV+0.2mm/mV. |
Auto record | Record setup according to auto record format and mode, automatically changing leads, measuring and a |
Enduring polarization voltage | +500mV |
Size | 334mm(L)x320mm(W)x85mm(H) |
Power supply/Frequency | AC: 100-240V,50/60Hz DC: 14.8V, 3700mAh rechargeable lithium battery |
Time constant | >=3.2s |
Measurement parameters | HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis |
CMRR | >60dB, >100dB( Adding filter) |
Input way | Floating and defibrillation protection |
Lead : | Standard 12 leads |
EMG interference filter | 25Hz/35Hz (-3dB) |
Paper speed | Auto record: 25mm/s, 50mm/s, error:5% Rhythm record: 25mm/s, 50mm/s, error:+5% Manual record: 5mm/s |
Noise level | <=15uVp-p |
Patient leak current | <10uA |
Product safety type | Class I CF applied part; there is defibrillation and pacing protection circuit in it. |